10m semi-anechoic chamber

10m semi-anechoic chamber


Dimensions of the test site 19.6(D) × 16.6(W) × 8.0(H) (m)

Measurement of emission
Test item Test port Specifications of main characteristics
Radiated interference signal Enclosure Measurement distance : 3m and 10m

3m semi-anechoic chamber

3m semi-anechoic chamber


Dimensions of the test site 7.0(D) × 4.0(W) × 3.0(H) (m)

Measurement of immunity
Test item Test port Specifications of main characteristics
Radiated immunity Enclosure Field strength : 10V/m (max)
Frequency range : 80 MHz ~ 2.7 GHz
Conducted immunity Power line
Signal line
Voltage : 10V (max)
Frequency range : 150 kHz ~ 80 MHz
CDN, capacitive clamp

Shield room

Shield room


Dimensions of test site 9.5(D) × 6.3(W) × 2.5(H) (m)

Measurement of immunity
Test item Test port Specifications of main characteristics
Conducted
interference
signal
Power line Measurement frequency range : 9 kHz ~ 30 MHz
AMN current capacity : single-phase:16A, three-phase:32A
Signal line Measurement frequency range : 9 kHz ~ 30 MHz
ISN, current probe

Shield room

shield room


SAR measurement
Test item Test port Specification of main characteristics
SAR measurement Head / Body Measurement frequency range : 450 MHz ~ 5.8 GHz

Please do not hesitate to contact us if you require advice or assistance.